www.labec.com.auLaboratory Equipment Pty Ltdemail: [email protected] Ph: 02 9560 2811 • Fax: 02 9560 61312014INSTRUCTION MANUAL
4.3 Open the door Press " " button of the control panel and the door cover will eject automatically to a certain height, then you need li
Don't use the rotor overload, or it will cause explosion of the rotor and the debris will damage the centrifuge. 4.7 Filling of sample in the
4.9.4 Time setting: Press the setting knob vertically on the control panel→make the number flash on the TIME →rotate the setting knob to set the speed
4.9.4 Time setting: Press the setting knob vertically on the control panel%make the number flash on the TIME %rotate the setting knob to set the speed
5th Items of Maintenance 5.1 Cleansing and purification If the hazardous materials spill out or into the device, users have the responsibility to do
5.1.3.4 Please wash the rotor with sponge or cotton bedewed with the neutral detergent, wash off the neutral detergent with distilled water. Don'
6. Troubleshooting 6.1 Open the cover in case of emergencies In the process of normal usage, sometimes there's accidental power failure or fail
Table 4: Fault phenomenon, causes and troubleshooting Fault phenomenon Causes and Troubleshooting No display 1. Check whether the power outlet and the
7. Packing List Packing list of Low-speed Table-top Centrifuge When you open the package, please check in time, if there's anything wr
8. Certificate of Compliance L-530 Low-speed Table-top Centrifuge Certificate of Compliance Maximum rotate speed: 5300r/min Maximum relative cent
USER MANUAL Low-speed Table-top Centrifuge L-530
Appendix 1 Optional Accessories Appendix 2 Technical data Rotor No. (Rotor) Capacity (ml×tubes) Maximum speed (r/m in) Maxim
Introduction The copyright of this manual belongs to Our company. Nobody is allowed to re-print or disseminate any information of the contents of thi
Safety Tips * Please read this manual carefully for the first-time using! * Low-speed Table-top Centrifuge can only be operated by trained or authori
Contents 1st Terms of safety using 5 1.1 Notes of Operation 6 2nd Introduction of Low-speed Table-top Centrifuge
1. Terms of safety using User’s better use it in accordance with the design of it, It may result in equipment damage or personal injury if there is
2. Introduction of Low-speed Table-top Centrifuge 2.1 Overview Low-speed Table-top Centrifuge can be widely used in fields of clinical medical, biolo
the rotor stops, you can press “ ” key of the control panel to open the door cover of the centrifuge. Only in the case of that the door cover of the c
and make the 4 supporting legs sustain evenly on the table-board, ensure that the mounting bracket and the table-board are fixed and don't swag.